女虐女变态耻辱调教小说_欧美aa一级片_青青草原国产在线观看_美女洗澡无遮挡物无内衣内裤

歡迎來到匠仁醫療設備有限公司網站!

更多 > 新聞資訊
公司新聞
行業知識
檢測答疑
圖標 首頁 >> 新聞資訊 > 行業知識

行業知識

內窺鏡外表皮及內管道破損、漏水的原因

來源:http://yuxiaous.cn/ 發布時間:2022-10-13 瀏覽量:0

內窺鏡是一種貴重的光學電子儀器,在內鏡取放、運輸、操作、洗消各個環節,均需要小心謹慎,要做到忙而不亂。下面內窺鏡維修公司來講講內窺鏡外表皮及內管道破損、漏水的原因。
Endoscope is a kind of valuable optical and electronic instrument. It needs to be careful in all aspects of endoscope taking, placing, transportation, operation and washing. Next, the endoscope maintenance company will talk about the causes of damage and leakage of the outer skin and inner pipe of the endoscope.
常見原因有從柜中取出內窺鏡時,內窺鏡與柜子四周發生碰撞。內窺鏡未按要求放在運輸車內盤好,內窺鏡打圈太小,先端部等被壓,加上運鏡時顛簸導致損壞。
The common reason is that when the endoscope is taken out of the cabinet, the endoscope collides with the cabinet. The endoscope was not placed in the transport vehicle as required, the circle of the endoscope was too small, the tip was pressed, and the endoscope was damaged due to turbulence during transport.
內窺鏡掛在主機上備用時,插入部擺動劇烈,與主機發生碰撞。術中沒有使用口墊,患者感到緊張口墊滑脫咬著插入部,內窺鏡與注射針等尖銳附件放一起致彎曲橡皮被劃破致漏水。
When the endoscope is hung on the host for standby, the insertion part swings violently and collides with the host. The mouth pad was not used during the operation. The patient felt that the mouth pad slipped and bit the insertion part. The endoscope and injection needle and other sharp accessories were put together, causing the bending rubber to be cut and water leakage.
即使是無痛胃鏡患者,操作期間突然蘇醒,如未及時護住口墊,也可導致內窺鏡被咬傷。洗消過程中破舊的刷子刷破管刷頭將內管道劃破。
Even if the patient with painless gastroscope suddenly wakes up during operation, if the mouth pad is not protected in time, the endoscope may be bitten. During the washing process, the worn brush will break the pipe and the brush head will scratch the inner pipe.
使用黏膜注射針及治療性附件不正確,插入時遇到阻力而強行將附件插入管道而導致鉗子管道漏水。
The use of mucosal injection needle and therapeutic accessories was incorrect, and the accessories were forcibly inserted into the pipe when encountering resistance, resulting in water leakage of the pliers pipe.
內窺鏡維修公司
處理措施:
Treatment measures:
從鏡柜中取出懸掛內窺鏡時,應當一手握住操作部,一手拿穩兩根軟管部,避免軟管呈銳角彎曲,對于激光共聚焦等復雜內窺鏡,應當兩人共同配合取放。放入運鏡車的內窺鏡,操作部、導光插頭部不能壓迫軟管,尤其是先端部。
When taking out the hanging endoscope from the mirror cabinet, hold the operating part with one hand and the two hose parts with the other hand to avoid sharp angle bending of the hose. For complex endoscopes such as laser confocal, two people should work together to take and place them. The operating part and light guide plug shall not compress the hose, especially the tip, when placing the endoscope of the mirror truck.
運鏡車一層一內窺鏡,禁止與其他內窺鏡或尖銳器械混放一起,運輸時下面墊一較厚的治療巾,避免運輸時顛簸損壞內窺鏡。
The transport vehicle shall have one endoscope on each floor. It is prohibited to mix with other endoscopes or sharp instruments. A thick treatment towel shall be placed under the transport vehicle to avoid damage to the endoscope due to turbulence during transport.
內窺鏡懸掛好后,插入部應當靜止下垂,避免左右擺動,本科內窺鏡主機與內窺鏡插入部可能接觸的部位均安裝了泡沫,從而緩沖內窺鏡與硬物的撞擊。
After the endoscope is suspended, the insertion part should be static and sagging to avoid swinging from left to right. Foam is installed at the parts that may contact the endoscope host and the insertion part of the endoscope, so as to cushion the impact between the endoscope and hard objects.
操作前,向患者耐心解釋,使其盡量放松;操作過程中,應當密切觀察、固定口墊,遇到緊急情況時,可提醒操作者退出內窺鏡,待患者平穩后再重新插入。
Before operation, patiently explain to the patient to make them relax as much as possible; During the operation, the mouth pad shall be closely observed and fixed. In case of emergency, the operator can be reminded to exit the endoscope and re insert it after the patient is stable.
經長時間的使用橡皮部會有磨損的痕跡,其原因是與牙墊摩擦有關,因此,操作者在進鏡、退鏡時應盡量減少橡皮與牙墊的接觸。內窺鏡禁止與尖銳器械放在同一個槽內洗消。為
After a long time of use, the rubber part will show signs of wear, which is related to the friction of the tooth pad. Therefore, the operator should try to reduce the contact between the rubber and the tooth pad when entering and withdrawing the mirror. Endoscopes shall not be put in the same groove with sharp instruments for washing. by
避免內窺鏡內管道損傷,應養成良好的操作習慣,操作前認真檢查附件性能,插送附件時要在順滑的條件下進行,不能用力太猛,遇到阻力時,應當松開角度扭,使內窺鏡插入部相對取直后再通過。更多相關注意事項就來我們網站http://yuxiaous.cn咨詢了解吧!
To avoid damage to the pipes in the endoscope, good operation habits should be developed. Before operation, carefully check the performance of the accessories. When inserting the accessories, they should be carried out in a smooth condition. Do not use too much force. When encountering resistance, loosen the angle torque to make the insertion part of the endoscope relatively straight before passing through. More relevant precautions come to our website http://yuxiaous.cn Ask and understand!
s
新聞資訊
聯系我們
  • 公司公司:匠仁醫療設備有限公司
  • 匠仁醫療山東濟南維修部
  • 聯系人:樊經理
  • 電話:13153199508
  • 地址:山東省濟南市槐蔭區美里東路3000號德邁國際信息產業園6號樓101-2室
  • 匠仁醫療湖南長沙維修部
  • 聯系人:李經理
  • 電話:13873135765
  • 地址:湖南省長沙市雨花區勞動東路820號恒大綠洲14棟2409室
新聞資訊

樊經理:13153199508     李經理:13873135765

公司地址:山東省濟南市槐蔭區美里東路3000號德邁國際信息產業園6號樓101-2室  湖南省長沙市雨花區勞動東路820號恒大綠洲14棟2409室

     

魯公網安備 37010402441281號