近幾年內窺鏡在我們的日常生活中應用越來越廣泛,但是許多人對于內窺鏡的知識并不是很清楚。實際上,內窺鏡泛指經各種管道進入人體,以觀察人體內部狀況的醫療儀器。那么內窺鏡設備在什么情況下容易被損壞?下面
內窺鏡設備維修人員為您講解吧。
In recent years, endoscope has become more and more widely used in our daily life, but many people are not very clear about the knowledge of endoscope. In fact, endoscope generally refers to medical instruments that enter the human body through various channels to observe the internal conditions of the human body. Under what circumstances are endoscope equipment vulnerable to damage? The endoscope equipment maintenance personnel will explain it to you.
使用中的注意事項硬管內窺鏡在手術過程中受到損壞的現象并不多,雖然會與人體的組織如肌肉、黏膜、骨骼等有接觸和磕碰,但是這些磕碰是輕微的,不會造成窺鏡的損壞,因為它只是起觀察的作用,不是其他器械的受力點。
Precautions in use: The rigid tube endoscope is not often damaged during surgery. Although it may contact and bump with human tissues such as muscles, mucous membranes, bones, etc., these bumps are slight and will not cause damage to the endoscope, because it is only for observation, not the force point of other instruments.
但是在使用其他器械時,尤其是咬合力較大的鉗、剪類器械應注意鏡管的前端不要伸進器械的咬合區內,以免誤傷鏡管。在使用這類器械時,有時醫生為了看清楚咬合區的組織,把窺鏡伸得很靠近組織,器械咬合時窺鏡沒有退回,誤傷了窺鏡。手術時如注意讓器械的咬合口全部都在窺鏡的觀察范圍內就可以避免此類事故發生。
However, when using other instruments, especially forceps and scissors with large bite force, the front end of the lens tube should not extend into the bite area of the instrument to avoid accidental injury to the lens tube. When using this kind of instrument, sometimes doctors stretch the endoscope very close to the tissue in order to see clearly the tissue in the occlusal area. When the instrument occludes, the endoscope does not return, and the endoscope is injured by mistake. During operation, such accidents can be avoided if all the occlusal openings of the instruments are within the scope of observation of the endoscope.
有些手術窺鏡是在鞘管內使用,在更換其他角度窺鏡或插拔器械時,應注意動作要輕,不可用力過猛。尤其是插拔窺鏡過程中,當遇到阻力拔不動時應仔細查找原因,必要時應連同鞘管一起拔取,不要用蠻力。
Some surgical scopes are used in the sheath tube. When replacing other angle scopes or inserting and pulling out instruments, attention should be paid to the light action and not to use too much force. Especially in the process of inserting and pulling out the endoscope, when it encounters resistance and cannot be pulled out, carefully find out the reason, and pull it out together with the sheath if necessary, without using brute force.
當窺鏡配合激光汽化、高頻電切、微波等光電技術進行手術時,應注意窺鏡前端與治療點的距離,保證窺鏡前端不被電擊或燒灼。當感覺到刨刀工作異常或照度突然降低時,有可能窺鏡已經受損,應及時更換,以免造成更大損失。
When the endoscope is operated with laser vaporization, high-frequency electrosection, microwave and other photoelectric technologies, attention should be paid to the distance between the front end of the endoscope and the treatment point to ensure that the front end of the endoscope is not electrocuted or burned. When you feel that the planer is working abnormally or the illumination is suddenly reduced, it is possible that the endoscope has been damaged, and you should replace it in time to avoid causing greater losses.
一般重要手術,應有一套備用窺鏡和關鍵器械,在發現問題時可以從容地更換;如果使用角度不合適的窺鏡或不配套的器械勉強手術,也容易造成窺鏡的損壞。了解了這些內容,希望您能了解,更多事項就來我們網站
http://yuxiaous.cn咨詢!
For general important surgery, there should be a set of spare endoscope and key instruments, which can be easily replaced when problems are found; It is also easy to cause damage to the endoscope if the endoscope with improper angle or mismatched instruments are used to operate reluctantly. After understanding these contents, I hope you can understand more and come to our website http://yuxiaous.cn consulting service