內(nèi)窺鏡分為硬鏡和軟鏡,硬鏡,一般使用在外科領(lǐng)域,插入管大多由不銹鋼材料組成,也有前部可彎曲的硬鏡,包括腹腔鏡、鼻竇鏡,前列腺電切鏡、輸尿管鏡及宮腔鏡等 。
Endoscope is divided into hard mirror and soft mirror, hard mirror, commonly used in surgical field, insert tube mostly by stainless steel material composition, also have front can bend hard mirror, include celioscope, nose sinuses mirror, prostate electrocision mirror, ureteroscope and hysteroscope to wait.
軟鏡,內(nèi)科領(lǐng)域比較多見,如腸胃鏡、十二指腸鏡、喉鏡、氣管鏡、插管鏡及輸尿管腎盂鏡等。
Soft endoscopy, medical fields more common, such as gastroscope, duodenoscope, laryngoscope, tracheoscope, intubation and Ureteropelvic Mirror and so on.
但是,在使用的過程中,內(nèi)窺鏡難免會出現(xiàn)這樣那樣的故障,一般它的故障出在哪兒,今天就讓匠仁醫(yī)療內(nèi)窺鏡設(shè)備維修的小團隊來給大家說一下。
However, in the use of the process, the endoscope will inevitably have this kind of failure, it is generally where the failure, today let artisan medical endoscope equipment repair team to tell you.
一般故障會出現(xiàn)以下幾個方面:
Common failures include the following:
顯示器沒有圖像顯示
為了確保圖象主機和顯示器已接通,圖象主機和顯示器的電源開關(guān)指示燈是否亮,如不亮,則需進行保險及供電檢查,檢查圖象主機與顯示器的連線、 IN和 OUT接口是否連接。在具有多種視頻信號源選擇器的顯示器中,選擇合適的視頻信號。查看顯示器的相應(yīng)設(shè)置參數(shù),將亮度、顏色、對比度默認設(shè)為默認,確保攝像機的連接電纜固定在圖像主機上。在圖象主機和顯示器之間更換視頻電纜,排除可能的電纜損壞。
Display no image display in order to ensure that the image host and the display are connected, whether the power switch indicator light of the image host and the display is on or not, if it is not, a power supply check should be carried out, check the connection between the image host and the Monitor, and the IN and OUT interfaces. In a display with a variety of video signal source selectors, select the appropriate video signal. Check the display settings, set the brightness, color, contrast to the default, to ensure that the camera cable fixed in the image host. Replace the video cable between the image host and the Monitor to eliminate possible cable damage.
監(jiān)視
錯誤的圖像顏色
錯誤的圖象顏色要求 WB重新校正。先確定 RGB或 YPbPr接口沒有接線錯誤,顯示參數(shù)確定,主機參數(shù)確認,選擇其它視頻通道進行測試,進行 COLOR BAR測試。檢查設(shè)定菜單的字體和顏色是否正常,如果正常,但內(nèi)視鏡圖像不正常,基本上是內(nèi)視鏡或攝像頭故障。
Monitor the wrong image color the wrong image color requires WB recalibration. First determine RGB or YPbPr interface no wiring error, display parameters determined, host parameters confirmed, select other video channels for testing, COLOR BAR testing. Check that the font and color of the settings menu are normal, if so, but the endoscopic image is not normal, basically an endoscopic or camera failure.
圖象干擾
發(fā)生干擾時應(yīng)檢查使用的高頻、強電的接地是否有問題,外部設(shè)備的信號線是否有問題。
If image interference occurs, check the grounding of the high frequency and high current used and the signal line of the peripheral equipment.
好了今天匠仁醫(yī)療
內(nèi)窺鏡設(shè)備維修給大家先介紹到這里,相信你看過后,以后會注意內(nèi)窺鏡的使用和維護了吧,接下來關(guān)于內(nèi)窺鏡的其它問題,關(guān)注官網(wǎng)
http://yuxiaous.cn/,我們繼續(xù)更新。
Okay, today, artisan medical endoscopic equipment maintenance to everyone first introduced here, I believe that after you have seen, you will pay attention to the use and maintenance of the endoscope, right? Next on the other issues of the endoscope, attention to the official website, we continue to update.