軟式內(nèi)鏡使用后應(yīng)當(dāng)立即用濕紗布擦去外表面污物,并反復(fù)送氣與送水至少10秒鐘;再用吸引通道反復(fù)吸清水或清洗液,直到流人吸引瓶的水干凈為止;后取下內(nèi)鏡,拆下各種附件,并裝好防水蓋,置合適的容器中送清洗消毒室。下面
內(nèi)窺鏡設(shè)備維修人員為您講解軟式內(nèi)鏡的水洗以及酶洗方式。
After using the flexible endoscope, immediately wipe the dirt on the external surface with wet gauze, and repeatedly send air and water for at least 10 seconds; Then repeatedly suck clean water or detergent through the suction channel until the water flowing into the suction bottle is clean; After that, take down the endoscope, remove various accessories, install the waterproof cover, and put it in a suitable container and send it to the cleaning and disinfection room. The endoscope equipment maintenance personnel will explain the water washing and enzyme washing methods of the flexible endoscope for you.
水洗
washing
(1)在流動(dòng)水下徹底沖洗,用紗布反復(fù)擦洗鏡身,同時(shí)將操作部清洗干凈;
(1) Rinse thoroughly under flowing water, scrub the mirror body repeatedly with gauze, and clean the operating part at the same time;
(2)取下活檢入口閥門、吸引器按鈕和送氣送水按鈕,用清潔毛刷徹底刷洗活檢孔道和導(dǎo)光軟管的吸引器管道,刷洗時(shí)必須兩頭見(jiàn)刷頭,并洗凈刷頭上的污物;
(2) Remove the biopsy inlet valve, aspirator button and air and water supply button, and thoroughly brush the aspirator pipe of the biopsy hole and light guide hose with a cleaning brush. When brushing, the brush head must be seen at both ends, and the dirt on the brush head must be cleaned;
(3)安裝全管道灌流器、管道插塞、防水帽和吸引器,用吸引器反復(fù)抽吸活檢孔道;
(3) Install the full-pipe perfusion device, pipe plug, waterproof cap and suction device, and use the suction device to repeatedly aspirate the biopsy hole;
(4)全管道灌流器接50毫升注射器,吸清水注入送氣送水管道;
(4) The whole pipeline perfusion device is connected with a 50ml syringe, and the water is sucked into the air supply and water supply pipeline;
(5)用吸引器吸干活檢孔道的水分并擦干鏡身。
(5) Use an aspirator to absorb water from the biopsy hole and dry the lens body.
將取下的吸引器按鈕、送水送氣按鈕和活檢入口閥用清水沖洗干凈并擦干。
Rinse and dry the removed aspirator button, water and air supply button and biopsy inlet valve with clean water.
內(nèi)鏡附件如活檢鉗、細(xì)胞刷、切開(kāi)刀、導(dǎo)絲、碎石器、網(wǎng)籃、造影導(dǎo)管、異物鉗等使用后,先放入清水中,用小刷刷洗鉗瓣內(nèi)面和關(guān)節(jié)處,清洗后擦干。
After the use of endoscopic accessories such as biopsy forceps, cell brushes, cutting knives, guide wires, lithotripters, net baskets, contrast catheters, foreign body forceps, etc., first put them into clean water, wash the inner surface of the clamp valve and joints with a small brush, and then wipe them dry.
清洗紗布應(yīng)當(dāng)采用一次性使用的方式,清洗刷應(yīng)當(dāng)一用一消毒。
The cleaning gauze should be disposable, and the cleaning brush should be disinfected once in use.
酶洗
Enzymolysis
(1)多酶洗液的配置和浸泡時(shí)間按照產(chǎn)品說(shuō)明書。
(1) The configuration and soaking time of multi-enzyme lotion shall be in accordance with the product manual.
(2)將擦干后的內(nèi)鏡置于酶洗槽中,用注射器抽吸多酶洗液100毫升,沖洗送氣送水管道,用吸引器將含酶洗液吸入活檢孔道,操作部多酶洗液擦拭。
(2) Place the dried endoscope in the enzyme washing tank, use a syringe to aspirate 100ml of multi-enzyme washing solution, flush the air supply and water supply pipeline, use an aspirator to inhale the enzyme-containing washing solution into the biopsy hole, and wipe the multi-enzyme washing solution of the operation department.
(3)擦干后的附件、各類按鈕和閥門用多酶洗液浸泡,附件還需在超聲清洗器內(nèi)清洗(5~10)分鐘。
(3) The dried accessories, various buttons and valves shall be soaked in multi-enzyme detergent, and the accessories shall be cleaned in the ultrasonic cleaner for (5~10) minutes.
(4)多酶洗液應(yīng)當(dāng)每清洗1條內(nèi)鏡后更換。
(4) The multi-enzyme lotion should be replaced after cleaning one endoscope.
軟式內(nèi)鏡的水洗以及酶洗方式相關(guān)內(nèi)容就講解到這里了,您對(duì)此有怎樣的疑惑或者需求就來(lái)我們網(wǎng)站
http://yuxiaous.cn咨詢!
That's all about the water washing and enzyme washing methods of flexible endoscope. If you have any doubts or needs about this, please come to our website http://yuxiaous.cn consulting service